|
BIOM 5100 / BMG 5103 / SYSC5302 / ELG6118B - Bioinstrumentation
Description
Instrumentation designed to measure physiological variables related to the function of the heart,lungs, kidney, nervous and musculo-skeletal system; emergency, critical care, surgery and anaesthesia equipment.
Prerequisites
OCIECE and OCIBME graduate students.
Instructor
Times and Locations
Fall 2009
(Sept. 14 − Dec. 7)
Google Calendar:
HTML
ICAL
Section | |
Activity | |
Day | |
Location | |
BIOM5100
| |
LEC 1
| |
Monday 8:30−11:30
| |
CO210
| |
Text
John G. Webster,
Medical Instrumentation: Application and Design,
− 4rd Edition:
ISBN: 0-471-67600-3 Hardcover 671 pages,
−
Text web site,
powerpoint slides
Marks
Work | |
Value |
Assignments
| |
24%
|
Project
| |
30−40%
|
− Project proposal
| |
− 3%
|
− Midterm report
| |
− 5%
|
− Presentation
| |
− 7%
|
− Report
| |
− 15%
|
Final Exam
| |
36−46%
|
Exams
Marks Policies
- Late work Policy (without *excellent* excuse):
1) 20% if ≤ 7 days late,
2) 0 mark if > 7 days late.
- If you wish to do a single (larger) assignments,
assignment for two graduate courses, you
must obtain written permission from
all profs involved. Otherwise, you may
find yourself in a position of academic fraud.
- If you have a question about a mark you have received,
fill out, sign and submit
this form.
- Academic fraud will be taken very seriously.
Cooperation between students for assignments is expected and
encouraged, however, copying of another's work
is not. You should not be leaving a discussion with
copies of another student's work.
Students with Disabilities
Students with disabilities requiring academic accommodations in this course must register with the Paul Menton Centre for Students with Disabilities for a formal evaluation of disability-related needs. Registered PMC students are required to contact the Centre, 613-520-6608, every term to ensure that I receive your Letter of Accommodation, no later than two weeks before the first assignment is due or the first in-class test/midterm requiring accommodations. If you require accommodation for your formally scheduled exam(s) in this course, please submit your request for accommodation to PMC by Nov 16, 2009.
Assignments
Assignments are due at the beginning of class
on the date indicated.
|
No. | |
Assignment | |
Due Date | |
1
| |
- Problem 2.1 (text).
Instead of "Ranges of Rm", use the
following values, Rm=100kΩ,
Rp=2kΩ
If you're having trouble with doing this question analytically,
you can solve it numerically with a simulation. I'll give
(almost) all the marks for this approach.
- Problem 4.6 (text).
Also, sketch the voltage distribution along the
nerve axon 10ms after the stimulation
Assume the nerve axon has no myelin and the
conduction velocity is 30m/s.
Sketch the regions of absolute and relative refractory
period
Note: you are only required to sketch a reasonable
shape of the curves. I don't require calculation
of the exact curve shape and values.
- Thermistors typically respond slowly. Consider a thermistor with
time constant τ=4.0s, and the subject to be breathing as
shown in Question 3C from
Final 07B.
Note: Only use the graph from this exam question.
You are not asked to solve the question on the
exam.
- The thermistor has
β=1000K; at T0=25°C,
R=250Ω.
What is the change in R for a change
from 40°C to 40.1°C?
- If we heat the thermistor to 40°C, we can measure the
cooling caused by the air flow. Assume that the breathing
flow, if kept constant, would cause a change of 3°C.
For an delay time of T=5.0s,
sketch the thermistor resistance vs time, starting
at an initial temperature of 40°C.
- Normally, a thermistor cannot measure the direction
of flow. Briefly explain why. However, with
two thermistors, we can measure flow by placing them
at different distances from the nostril.
Briefly explain why the signals in the thermistors
will differ. Sketch the signals in each thermistor.
- Discuss one biomedical application of a
(1) Piezoelectric sensor, and
(2) strain gauge sensor.
(each paragraph should be approx 100 words. This question
requires some research beyond the course material.)
Note: be careful to avoid plagiarism if you quote from
source on the internet (note that I'm not suggesting
that you may not use the internet)
| |
Oct. 5
| |
|
2
| |
- (5 points) Often the electrolyte gel on an Ag/AgCl begins to dry out as
after prolonged use (24h+). Draw the equivalent circuit
for an electrode & gel. Discuss
how this circuit
changes when the electrode gel drys out.
- (5 points) Question 1B from Final 07B
- (5 points) Question 1A from Final 08
- (5 points) Question 1B from Final 08
- (5 points) Question 1F from Final 08
(only the first question change in electrode voltage)
- (5 points) In class, we discussed some mechanisms of ventricular
fibrillation. Discuss how a relatively long absolute
refractory period can help protect against ventricular
fibrillation.
- (5 points) Nerve Conduction Velocity:
Consider the classic "knee-jerk" reflex test.
Sketch a diagram of the nerve signal
conduction pathway during this test.
What accounts for the time delay between the hammer tap
and muscle contraction?
| |
Oct. 26
| |
|
3
| |
-
Sketch a graph of blood pressure, ventricular volume, heart sounds and
electrical activity (slide #14.7).
Sketch and briefly describe how the graph of pressure and volume
would change if the aortic valve were
a) Stenotic,
b) Insufficient
- Consider two sources of interference on an
an instrumentation amplifier: common-mode
interference and thermal noise in a resistor.
An INA 128
instrumentation amplifier with CMRR=120dB is used to measure signals:
Vcm=10V and
Vd=100µV.
-
If RG=500Ω what is the output, Vo
for the difference and common-mode part of the signal.
- If the resistor just above op-amp A3 (on page 1 of the specification shee)
has a value of 40.1kΩ (instead of 40kΩ), and the chip is
otherwise perfect, calculate the CMRR.
- In practice, the cables that are used to connect the amplifier
to the patient mean that much lower CMRR values are obtained.
Discuss (briefly) one reason for this decrease in CMRR
- Based on Final 2005 (Q3).
An intra-arterial blood pressure transducer is to measure
systolic and diastolic pressure. Assume the blood pressure has
simplified rectangular waveform, as shown. The systolic pressure
is 150 mmHg; the diastolic pressure is 100 mmHg. Systole
lasts 200 ms, while diastole lasts 500 ms.
- In class we descussed three kinds of pressure waveform distortions.
Briefly describe (<50 words) each of them, using a sketch.
-
Assume, the pressure transducer is responding very slowly. Measurement
of the response shows f0=20 Hz, and ζ=1.5.
a) Sketch the blood pressure waveform and the blood pressure transducer output.
b) By how much will the pressure underestimate the true values?
-
Now, assume, the pressure transducer has
f0=20 Hz, and ζ=0.15.
a) Extimate the fraction of overshoot (slide 2A.16).
a) Sketch the blood pressure waveform and the blood pressure transducer output.
b) By how much will the pressure overestimate the true values?
-
A person with a systolic pressure of 150 mmHg and
diastolic pressure of 90 mmHg and HR=90 beats/min,
is having their blood pressure measured.
-
Using the oscillometric technique, sketch a graph of the blood pressure
versus time.
(Some people have put only the cuff pressure or the blood pressure.
I'll accept both).
Indicate what is happening at each significant point (Systolic Pressure,
Mean Arterial Pressure, and Diastolic Pressure).
-
It it possible to obtain a continuous monitoring of blood pressure
using the oscillometric technique?
Briefly describe why or why not
| |
Nov. 9
| |
|
4
| |
- A pulse oxymeter is used to measure arterial oxygen
concentration. However, arterial blood represents a small fraction of the
material in the finger or ear lobe onto which the oxymeter is typically
placed. Most of the optical path is filled with skin, tissue and venous
blood. Explain how a pulse oxymeter is able to measure arterial blood
concentration alone and to reject the contribution from other tissues.
- If a pulse oxymeter uses only two wavelengths of light, it
can be fooled by the presence of HbCO (carboxyhemoglobin) in the blood.
How would this change for a pulse oxymeter which uses
more light wavelengths?
- Do the following questions from the
2008 final exam
- Question 2A.
- Question 2B.
- Question 2C. (Answer Both (1) and (2) subsections)
- Question 2D.
| |
Nov. 23
| |
Project
Activity
| |
Description
| |
Due Date
| |
Project Proposal
| |
Proposal is 1-2 pages (double spaced). Include problem
description, proposed techniques, and references.
| |
Oct. 5
| |
Midterm report
| |
Report is a minimum of 5 pages (double spaced). Describe
project progress to date. Discuss any current challenges.
| |
Nov. 16
| |
Presentation
| |
Presentations will be (≤10 minutes) in English.
Marks are based on technical content (45%), clarity of
presentation (45%), and ability to answer questions (10%).
| |
Nov. 30 − Dec. 7
| |
Project Report
| |
The report will be 10−20 pages (double spaced) and written in
English or French. Include an introduction, methods, results and
discussion. Ensure that the report synthesizes and
discusses the material. Do not simply restate
existing work.
Marks are based on the technical content, and clarity of its
presentation.
| |
Dec. 14
| |
The Course project may be any of the following:
( it will be permissible to work in groups of 2 or 3 for
projects in which an experiment is designed - the larger
the group, the more challenging the project).
- Design of a medical instrumentation experiment.
A list of possible project ideas
(please discuss your ideas with the instructor
to get feedback on feasibility)
− measure ECG and calculate heart rate for different activities
− measure EMG and calculate gait
− measure EMG and calculate fatigue
− measure flow and calculate Flow Volume loops to get maximum
expiratory flow
− build mask to simulate high expiratory resistance in
asthma − calculate flow
− get blood pressure from inflated cuff
− Measure VO2,max with exercise
− Test maximum expiratory flow with negative end
expiratory pressure (NEEP)
− Measure surface and core temperature changes
- Implementation and testing of a medical instrumentation
algorithm
Data may come from any of the following sources:
− Measured at respiratory system lab
− Obtained from a biomedical database
- In depth study of a medical instrumentation
application
A project of this type must provide an in depth
investigation of the scientific literature for
a specific application. Students must ensure
their report is comprehensive and
synthesizes information from multiple
sources. Reports which simply follow one primary
reference will not be worth high marks.
Course Outline
Date
| |
Activity
| |
Sept. 14
| |
Introduction,
Measurement systems (ch 1)
−
Slides #1,
Overview of basic sensors (ch 2)
−
Slides #2,
Slides #2a,
Slides #3
|
Sept. 21
Oct. 5
| |
Overview of electrical safety (ch 14)
−
Slides #4,
Bioelectrical signals (EMG, ECG, EEG) (ch 4)
−
Slides #5,
Slides #6,
Slides #7,
Slides #8
|
Sept. 28
| |
Biomed Signals Lab. (Minto 6070)
− Use of CleveMed equipment:
− Equipement: pneumotach, electrode amplifiers, force, pulse oxymeter,
blood pressure, stethoscope
− Use of NI amplifiers, filtering, 60Hz notch, AC vs DC preamp
− Measurements: ECG + exercise, EMG with reflex response,
Labs from undergrad instrumentation course
(username is "sysc4203" passwd is same):
ECG,
EMG.
|
Oct. 12
| |
Thanksgiving holiday (no class)
|
Oct. 19
| |
Electrodes (ch 5)
−
Slides #9,
Biopotential Amplifiers (ch 3,6)
−
Slides #10,
Slides #11,
Slides #12,
AD620 [Analog Devices]
INA128 [Texas Instruments]
|
Oct. 26
Nov. 2
| |
Measurement of blood pressure and sound (ch 7)
−
Slides #13,
Slides #14,
Slides #15,
Slides #16,
Slides #17,
Slides #18,
|
Nov. 9
| |
Measurement of blood flow, volume (ch 8);
Photoplethysmography and pulse oxymetry
−
Slides #19,
Slides #20,
Slides #21,
Slides #22,
|
Nov. 16
Nov. 23
| |
Measurement of Respiratory system (ch 9)
−
Slides #23,
Slides #24,
Slides #25,
Slides #26,
Slides #27,
|
Nov. 30
Dec. 7
| |
QRS Detection: Slides #28,
Electrical Impedance Tomography,
Project Presentations
|
Dec. 11
| |
Review Class: 10h00−12h00: CO210,
Project reports due:
|
Dec. 14
| |
Final Exam: 13h00−16h00 in LA A620
|
Last Updated:
$Date: 2023/01/10 14:28:58 $
|